Designed for the professional laboratory, Aircare’s Labrocare Fume Cupboard provides proven fume extraction technology, constant air volume (CAV), a wide range of sizes, and optional special features for demanding environments. The Labrocare Fume Cupboard fully complies with AS/NZS 2243.8 (2014) standards.
Constant Air Volume fume cupboards have their exhaust fan set to a fixed speed, calculated to meet the minimum airflow demanded by the relevant standards when the sash is at normal working height. Advanced Air® fume cupboards ensure 0.5 metres/second airflow when the sash is at the recommended working height (normally 500 mm).
To prevent excessive air velocities through the sash opening when the sash is lowered, Labrocare CAV fume cupboards use an Air Bypass System. At the normal working height, incoming air flows in exclusively through the sash opening. As the sash is lowered, part of the airflow is drawn in through the bypass inlet on top of the fume cupboard.
The bypass prevents excessive velocity at the sash opening by providing a balanced airflow into the fume cupboard. If the sash is opened above normal working height – 730 mm is maximum – airflow at the sash will be reduced to an unsafe level. Labrocare fume cupboards provide a lock at normal working height. It serves as a barrier that must be disabled by the operator to raise the sash above 500 mm; it automatically relocks when the sash is lowered.